Pro-CT AAPM


The Pro-CT AAPM phantom may be used for carrying out tests of computed tomography systems according to Report no. 1 of the AAPM (American Association of Physicists in Medicine).


​​for more information Click Here

Computer Tomography

Pro-CT Dose AAPM TG-200 / Pro-CT Dose ICRU 87


​This phantom is designed according to ICRU 87 report and AAPM TG-200 specifications.
The phantom comprises high-density polyethylene (mass density of 0.97 g/cm3) and is 300mm in diameter. The total length of the phantom is 600mm. This allows to capture the majority of the scattered radiation and gives more accurate results than standard PMMA cylinders – 150mm long, 320mm in diameter.
Due to the large mass of the phantom, it is manufactured in three separate sections. In order to keep these sections aligned, alignment pegs are included in the design.


​​for more information Click Here

Pro-CT Dose AAPM TG-111


​This phantom is designed according to AAPM TG-111 findings. The phantom comprises three 20cm thick PMMA cylinders 300 mm in diameter similar to standard Pro-CT Dose phantom. The total length of the phantom is 600 mm.This allows to capture the majority of the scattered radiation and gives more accurate results than standard PMMA cylinders – 150mm long, 320mm in diameter.
Due to the large mass of the phantom, it is manufactured in three separate sections. In order to keep these sections aligned, mounting screws are included in the design.


​​for more information Click Here

​Pro-CT Dose


Set of homogenous cylinders designed to image paediatric and adult head and body.
Used for CTDI measurement and to separate dose information for each. When performing dose profile measurements, dose phantoms allow to collect information for the maximum, minimum and mid-range value of the nominal tomographic section thickness.


​​for more information Click Here

Pro-CT


​The Pro-CT phantom may be used for carrying out acceptance and constancy tests of computed tomographysystems according to IEC 61223-3-5 and the AAPM (American Association of Physicists  in Medicine) guidelines.
This phantom consists of five test modules placed on an adjustable stand allowing for accurate alignment on the CT table. Markings on modules provide further facilitation of the positioning process. There is no need to fill the phantom with water, which can be very inconvenient, because all modules are filled with a substance whose density is similar to water.


​​for more information Click Here

Pro-CT Align


This phantom can be used to check the alignment of the internal and external lasers to the radiographic center of CT and PET/CT units and to verify lateral gantry angle. It can also be used with accelerator units to check vertical and lateral gantry angles, laser alignment and vertical table movement.


​​for more information Click Here